INCUBATOR 30l (INCUDIGIT-SV 30L), 230V 50-60Hz

STD ELAEINCE10-

Valid Article

CE marking: declaration that the product meets EU standards for health, safety, and environmental protection. The CE marking indicates that the product may be sold freely in any part of the European Economic Area, regardless of its country of origin.
The order of this product needs to be justified and is only acceptable under certain conditions.
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

See Related

Add article to list

   

Add Kit To Wishlist

   

INCUBATOR 30l (INCUDIGIT-SV 30L), 230V 50-60Hz

Definition

Device for monitoring and controlling the temperature and the humidity to provide ideal conditions for the growth of microorganisms.

This model contains extra isolation. This gives an autonomy of between 8 and 12 hours in the event of a power cut.

Benchtop model with one door

Specifications

Some restricted information has been hidden. Sign in to see this information

Technical specifications

  • Biological incubator isolated using phase change material. This gives it an autonomy of between 8 and 12 hours in the event of a power cut.
  • Rated power: 150W

Dimensions

  • Internal volume: 30L
  • Interior: h 35 / L 31 / d 29 cm
  • Exterior: h 69 / L 56 / d 57 cm
  • Weight: 52 kg

Instructions for use

Read carefully the supplier 'Instruction Manual' before first use.

Precautions for Use

Secure the electrical protection of the device with a double conversion UPS (see related articles below).

Maintenance

A monthly cleaning of the inside of the incubators and shelves must be carried out using absorbent paper soaked in dilute Surfanios.

Note that this cleaning can and should be carried out more often if necessary (for example in case of suspicion of contamination or projections).

  • Set the incubator to OFF;
  • Take the content out of the incubator;
  • Remove the shelves;
  • Clean the inside of the incubator and the shelves and replace them;
  • Put the incubator on ON;
  • Return the contents to the incubator

MSF requirements

Reserved for projects using the Mini-Lab - Optional in case of deployment on a site particularly sensitive to power cuts